Caesium iodide (CsI), with thallium and sodium doped scintillator grades CsI
CsI · CAS 7789-17-5 · cubic, Pm3m, CsCl structure
Caesium iodide is the cubic CsCl-structure crystal CsI, grown by the sealed ampoule Stockbarger method in ingots of about 70 mm diameter.
Evidence level
2 of 5
Lab-measured
- 0Predicted only. No one has made it, or no measurement exists.
- 1Calculated. Only calculated or predicted values from sources.
- 2Lab-measured. Measured properties with their conditions stated.
- 3Space-relevant tests. Measured in conditions relevant to space, such as vacuum or atomic oxygen.
- 4Ground-qualified. Qualified for an application on the ground.
- 5Flown or field-proven. Flight or field results exist.
Range
0.25 to 55 µm
Refraction
1.73916
Light
550 nm
Heat
621 °C
Density
4.51 g/cc
Recipe, share of the weight
- Cs not statedCaesium
- I not statedIodine
- Tl not statedThallium
- Na not statedSodium
Best for
IR windows that must reach far into the infrared, and X-ray and gamma scintillators read by silicon photodiodes or SiPMs (CsI(Tl))
Weakest at
Softness and moisture: the crystal is extremely soft and hard to polish, and it is slightly hygroscopic and soluble in water
Risk at a glance
- People Low. Irritant; may cause allergy or breathing irritation
- If it fails High. Dissolution in water
- Corrosion Severe. CsI dissolves in water, at 44 g per 100 g of water at 0 °C.
Quick facts: CsI
| Property | Value | Basis |
|---|---|---|
| Density | 4.51 g/cm³ | 1 sourced value |
| Young's modulus | 5.3 GPa | 1 sourced value |
| Thermal conductivity | 1.1 W/(m·K) | 1 sourced value |
| Thermal expansion (CTE) | 48.3 to 51 ppm/K | 3 sourced values across conditions |
| Melting point | 621 °C | 1 sourced value |
| Specific heat | 201 J/(kg·K) | 1 sourced value |
| Maximum service temperature | 125 °C | 1 sourced value |
What is the density of Caesium iodide (CsI)?
Caesium iodide (CsI) has a density of 4.51 g/cm³. The source and test conditions are listed in the properties table on this page.
What is the thermal conductivity of Caesium iodide (CsI)?
Caesium iodide (CsI) has a thermal conductivity of 1.1 W/(m·K). The source and test conditions are listed in the properties table on this page.
What is the coefficient of thermal expansion of Caesium iodide (CsI)?
Caesium iodide (CsI) has a coefficient of thermal expansion of 48.3 to 51 ppm/K. The range covers 3 sourced values across the tempers, forms and test conditions the sources report; room-temperature values are used where a source states a temperature.
What is the melting point of Caesium iodide (CsI)?
Caesium iodide (CsI) has a melting point of 621 °C. The source and test conditions are listed in the properties table on this page.
What is the Young's modulus of Caesium iodide (CsI)?
Caesium iodide (CsI) has a Young's modulus of 5.3 GPa. The source and test conditions are listed in the properties table on this page.
What is Caesium iodide (CsI) used for?
Uses recorded in the sources include IR spectrophotometer components; X-ray image intensifier input phosphor and X-ray panels; Gamma-ray telescope calorimeter (Fermi LAT).
Why it exists
Caesium iodide is a cubic crystal that transmits from 0.25 to 55 µm (Crystran), so it is used where the infrared reaches furthest. Doped with thallium it becomes a 550 nm scintillator that silicon photodiodes read (BNC). Doped with sodium it becomes a 420 nm scintillator that bialkali photomultipliers read (Shalom EO). The Fermi LAT calorimeter flew CsI(Tl) in low Earth orbit and measured a light yield loss of about 1 percent a year.
Where it's used
IR spectrophotometers
Crystran says CsI is sometimes used for components in the widest range spectrophotometers, because it has the deepest known IR transmission.
X-ray panels and image intensifiers
USGS says CsI is used primarily in X-ray panel production, and as the input phosphor of X-ray image intensifier tubes in fluoroscopy.
Gamma-ray telescope (Fermi LAT)
The LAT calorimeter has 8.6 radiation lengths of CsI(Tl) crystals in 8 layers, and its light yield is tracked in orbit.
Recipe
CsI
Caesium iodide is an ionic compound of caesium and iodine, with one caesium atom for each iodide ion. Crystran grows the crystal by the sealed ampoule Stockbarger technique in ingots of about 70 mm diameter. Thallium or sodium is added as a dopant to make the scintillator grades, CsI(Tl) at 0.10 to 0.12 mole percent thallium and CsI(Na) with sodium.
| Ingredient | Share of the weight | What it does |
|---|---|---|
| Caesium Cs | not stated | The alkali metal cation. One caesium atom per formula unit (PubChem CID 5354618 for the atomic weight). |
| Iodine I | not stated | The halogen anion. One iodine atom per formula unit (formula CsI from PubChem CID 24601). |
| Thallium Tl | not stated | Activator dopant in the CsI(Tl) grade, at 0.10 to 0.12 mole percent (BNC). |
| Sodium Na | not stated | Activator dopant in the CsI(Na) grade (Shalom EO). |
Weight shares are as printed in the sources.
What it can do
Range
0.25 to 55 µm
Crystran's data table gives a transmission range of 0.25 to 55 µm for undoped CsI, the widest of the windows listed on the page.
Refraction
1.73916
Refractive index 1.73916 at 10 µm, per the Crystran data table. Refractive index 1.79 at the CsI(Tl) emission maximum (BNC).
Light
550 nm
CsI(Tl) emits at 550 nm (BNC) and CsI(Na) emits at 420 nm (Shalom EO), so each grade suits a different photosensor.
Heat
621 °C
Melting point 621 °C on the Crystran table. BNC gives a useful temperature range up to 125 °C for CsI(Tl).
Density
4.51 g/cc
Density 4.51 g/cc for the crystal, the same value printed for CsI(Tl) and CsI(Na).
How it compares
Crystran's sheet and the two scintillator vendors give different figures for each grade, so the comparison is between grades and not between suppliers. Undoped CsI transmits to 55 µm, and its thermal conductivity is 1.1 W/(m·K) at 298 K on the Crystran table, a value higher than the 0.13 that BNC gives for CsI(Tl). CsI(Tl) is about 51 ppm/K in thermal expansion on the BNC sheet, and CsI(Na) is 49 on the Shalom EO chart. The sheets give no side-by-side ground or space test and no cost comparison.
The same measurements for each material, so you can compare them. Each row names the key condition under the value; the full conditions and methods are in the numbered notes under Technical details. Other materials are shown beside it; each was measured under its own conditions, so open its page before comparing closely. Compare every grade in the family.
| Measurement | CsI | Al2O3 | Cr:Al2O3 | CaF2 | LiF | ZnSe | ZnS | CdSe | Be3Al2Si6O18 / SiO2 / CuAl6(PO4)4(OH)8 / KAlSi3O8 / (Mg,Fe)2SiO4 |
|---|---|---|---|---|---|---|---|---|---|
| Thermal behavior and safety | |||||||||
| Melting point | 621 °C6 | 2040 to 2054 °C | 2040 °C | 1360 °C | 848 °C | 1525 °C | 1827 °C | 1537 °C | no data |
| Thermal conductivity | 1.1 W/(m·K)8 | 42 W/(m·K) | 0.35 W/(cm·K) | 9.71 W/(m·K) | 11.3 W/(m·K) | 18 W m-1 K-1 | 27.2 W/(m·K) | 6.9 W/(m × K) | no data |
| Specific heat capacity | 201 J/(kg·K)12 | 0.75 J/(g·K) | 761 J/(kg·K) | 854 J/(kg·K) | 1562 J/(kg·K) | 0.339 J/(g K) | 515 J/(kg·K) | 258 J/(kg × K) | no data |
| Physical and electrical | |||||||||
| Density | 4.51 g/cc5 | 3.97 g/cm³ | 3.98 g/cm³ | 3.18 g/cc | 2.639 g/cc | 5.27 g/cm³ | 4.09 g/cm³ | 5.81 g/cm³ | 2.63 to 2.97 g/cm³ |
| Temperature behavior | |||||||||
| Refractive index change with temperature (dn/dT) | -99.3 ppm/K2 | no data | no data | -10.6 ppm/K | -16 ppm/K | no data | no data | no data | no data |
| Other measurements | |||||||||
| Refractive index at 10 µm | 1.739 -1 | 1.768 - | 1.759 - | 1.399 - | 1.392 - | 2.403 - | 2.201 - | 2.431 - | 1.567 to 1.61 - |
| Refractive index at the emission maximum, CsI(Tl) | 1.79 -3 | 1.768 - | 1.759 - | 1.399 - | 1.392 - | 2.403 - | 2.201 - | 2.431 - | 1.567 to 1.61 - |
| Refractive index at the emission peak, CsI(Na) | 1.84 -4 | 1.768 - | 1.759 - | 1.399 - | 1.392 - | 2.403 - | 2.201 - | 2.431 - | 1.567 to 1.61 - |
| Maximum useful temperature, CsI(Tl) crystal | 125 °C7 | no data | no data | no data | no data | 250 °C | 250 °C | no data | no data |
| Coefficient of thermal expansion | 48.3 ppm/K9 | 7.7 ppm/K | 4.78 ppm/K | 18.85 ppm/K | 37 ppm/K | 7.1 x10-6/K | 6.5 ppm/K | 2.45 x10-6/K | no data |
| Coefficient of thermal expansion, CsI(Tl) | 51 ppm/K10 | 7.7 ppm/K | 4.78 ppm/K | 18.85 ppm/K | 37 ppm/K | 7.1 x10-6/K | 6.5 ppm/K | 2.45 x10-6/K | no data |
| Coefficient of thermal expansion, CsI(Na) | 49 ppm/K11 | 7.7 ppm/K | 4.78 ppm/K | 18.85 ppm/K | 37 ppm/K | 7.1 x10-6/K | 6.5 ppm/K | 2.45 x10-6/K | no data |
| Dielectric constant at 1 MHzValue printed on the Crystran product data table | 5.65 - | 11.5 - | no data | 6.76 - | no data | no data | no data | 9.7 - | no data |
| Young's modulus | 5.3 GPa13 | 470 GPa | no data | 75.8 GPa | 64.97 GPa | 67.2 GPa | 74.5 GPa | no data | no data |
| Shear modulusValue printed on the Crystran product data table | 6.24 GPa | 148.1 GPa | no data | 33.77 GPa | 55.14 GPa | no data | no data | no data | no data |
| Bulk modulusValue printed on the Crystran product data table | 12.67 GPa | 240 GPa | no data | 82.71 GPa | 62.03 GPa | 40 GPa | no data | no data | no data |
| Poisson ratioValue printed on the Crystran product data table | 0.214 - | 0.18 to 0.29 - | no data | 0.26 - | 0.326 - | 0.28 - | 0.28 - | no data | no data |
| Apparent elastic limit | 5.6 MPa14 | 300 MPa | no data | 36.54 MPa | 11.2 MPa | no data | 26.9 MPa | no data | no data |
| Reststrahlen peak | 145.8 µm15 | no data | no data | 35 µm | 25 µm | no data | no data | no data | no data |
| Transmission range | 0.25 to 55 µm16 | no data | no data | 0.13 to 10 µm | 0.12 to 6 µm | 0.6 to 21 µm | no data | 0.75 to 25.5 µm | no data |
| Solubility in water at 0 °C | 44 g/100 g water17 | 9.8×10^-5 g/100 g water | no data | 0.0017 g/100 g water | 0.27 g/100 g water | 0.001 g/100 g water | 6.5×10^-5 g/100 g water | no data | no data |
| Emission maximum, CsI(Tl) | 550 nm18 | no data | no data | no data | no data | no data | no data | no data | no data |
| Emission peak, CsI(Na) | 420 nm19 | no data | no data | no data | no data | no data | no data | no data | no data |
| Decay time, fast component, CsI(Tl) | 600 ns20 | no data | no data | no data | no data | no data | no data | no data | no data |
| Decay time, slow component, CsI(Tl) | 3.4 µs21 | no data | no data | no data | no data | no data | no data | no data | no data |
| Decay time, CsI(Na) | 630 ns22 | no data | no data | no data | no data | no data | no data | no data | no data |
| Photoelectron yield with a bialkali PMT, CsI(Tl) | 45 % of NaI(Tl)23 | no data | no data | no data | no data | no data | no data | no data | no data |
| Photoelectron yield, CsI(Na) | 85 % of NaI(Tl)24 | no data | no data | no data | no data | no data | no data | no data | no data |
| Lower wavelength cutoff, CsI(Tl) | 320 nm25 | no data | no data | no data | no data | no data | no data | no data | no data |
| Refractive index at the emission maximum, undoped CsI | 1.95 -26 | 1.768 - | 1.759 - | 1.399 - | 1.392 - | 2.403 - | 2.201 - | 2.431 - | 1.567 to 1.61 - |
| Thallium doping concentration, CsI(Tl) | 0.1 to 0.12 mole%27 | no data | no data | no data | no data | no data | no data | no data | no data |
Why choose it, and why not
Good for
- IR windows and spectrophotometer components that must transmit to 55 µm (Crystran).
- X-ray panels and image intensifier input phosphors (USGS).
- CsI(Tl) scintillators read by silicon photodiodes or SiPMs (BNC).
Watch out for
- Humidity. CsI is slightly hygroscopic (CsI(Tl)) and hygroscopic (CsI(Na) per Shalom EO). Keep bare crystals dry.
- Softness. The crystal deforms and is hard to polish, so the transmission quality is compromised for range.
- Thermal ramps. The high thermal expansion means slow ramp rates are needed for larger crystals (BNC).
- Orbit. The Fermi LAT calorimeter lost about 1 percent a year of light yield, and a correction is applied.
- Source conflicts. Vendors disagree on thermal conductivity, thermal expansion and decay time (see gaps).
Caesium iodide is a soft, moisture-sensitive crystal with the widest IR window of the common halides, and as CsI(Tl) it is a rugged scintillator that a silicon photodiode can read. The sources read describe its optics and its one flight use, but give no material test in space, so its orbital life is not established by this profile.
Technical details
For scientists and researchers. Everything above comes from the record below: every number with its method, its conditions and its source.
Notes
The method, full conditions and any caveat for the numbered rows above.
| # | Measurement | Method | Conditions | Caveat |
|---|---|---|---|---|
| 1 | Refractive index at 10 µm | Printed on the Crystran product data table, with references to the Palik handbook and a 1955 paper | wavelength 10 µmdopant none (undoped crystal) | — |
| 2 | Refractive index change with temperature (dn/dT) | Printed as -99.3 x 10^-6 K^-1 with reference 2 (Rodney 1955); the number is not converted; not a canonical property name | unit in source 10^-6 K^-1wavelength not stated on the pagedopant none (undoped crystal) | — |
| 3 | Refractive index at the emission maximum, CsI(Tl) | Printed on the Berkeley Nucleonics CsI(Tl) datasheet (2016), specifications table | wavelength 550 nm (emission maximum)dopant thallium, 0.10 to 0.12 mole% | — |
| 4 | Refractive index at the emission peak, CsI(Na) | Printed on the Shalom EO CsI(Na) product page, specifications table and comparison chart | wavelength 420 nm (emission peak)dopant sodium | — |
| 5 | Density | Value printed on the Crystran product data table; the same 4.51 g/cm3 is printed for CsI(Tl) and CsI(Na) on the BNC and Shalom EO pages | temperature not stated on the pagedopant none (undoped crystal) | — |
| 6 | Melting point | Value printed on the Crystran product data table; the Shalom EO comparison chart prints 621 °C for CsI, CsI(Tl) and CsI(Na) | pressure not stated on the pagedopant none (undoped crystal) | — |
| 7 | Maximum useful temperature, CsI(Tl) crystal | Printed on the BNC CsI(Tl) datasheet as the upper end of a useful temperature range; the lower end is left out (see gaps) | dopant thallium, 0.10 to 0.12 mole%basis bare crystal, useful temperature range | — |
| 8 | Thermal conductivity | Printed on the Crystran product data table with reference 3 (Combes et al. 1951); the BNC datasheet prints 0.13 for CsI(Tl), which is not reconciled here (see gaps) | temperature 298 Kdopant none (undoped crystal) | — |
| 9 | Coefficient of thermal expansion | Printed as 48.3 x 10^-6 K^-1 at 293 K; the number is not converted | unit in source 10^-6 K^-1temperature 293 Kdopant none (undoped crystal) | — |
| 10 | Coefficient of thermal expansion, CsI(Tl) | Printed on the BNC CsI(Tl) datasheet as 51 x 10^-6 K^-1; the number is not converted; the Shalom EO chart gives 54 for CsI(Tl) (see gaps) | unit in source 10^-6 K^-1temperature not stated on the pagedopant thallium, 0.10 to 0.12 mole% | — |
| 11 | Coefficient of thermal expansion, CsI(Na) | Printed on the Shalom EO CsI(Na) page as 49x10^-6 K^-1; the number is not converted | unit in source 10^-6 K^-1temperature not stated on the pagedopant sodium | — |
| 12 | Specific heat capacity | Printed on the Crystran product data table with reference 4 (Kelly 1934); the BNC datasheet prints the same 201 J/kg K | temperature not stated on the pagedopant none (undoped crystal) | — |
| 13 | Young's modulus | Value printed on the Crystran product data table; the BNC datasheet prints 5.3 GPa for CsI(Tl) | temperature not stated on the pagedopant none (undoped crystal) | — |
| 14 | Apparent elastic limit | Printed on the Crystran product data table as 5.6 MPa (810 psi); not a canonical property name | temperature not stated on the pagedopant none (undoped crystal) | — |
| 15 | Reststrahlen peak | Printed on the Crystran product data table; not a canonical property name | temperature not stated on the pagedopant none (undoped crystal) | — |
| 16 | Transmission range | Range printed on the Crystran product data table; not a canonical property name | thickness not stated on the pagedopant none (undoped crystal) | — |
| 17 | Solubility in water at 0 °C | Printed on the Crystran product data table; not a canonical property name | temperature 0 °Cdopant none (undoped crystal) | — |
| 18 | Emission maximum, CsI(Tl) | Printed as the maximum emission wavelength on the BNC CsI(Tl) datasheet; the Shalom EO comparison chart also prints 550; not a canonical property name | dopant thallium, 0.10 to 0.12 mole%temperature not stated on the page | — |
| 19 | Emission peak, CsI(Na) | Printed on the Shalom EO CsI(Na) page as the emission peak wavelength; not a canonical property name | dopant sodiumtemperature not stated on the page | — |
| 20 | Decay time, fast component, CsI(Tl) | Printed on the BNC CsI(Tl) datasheet as 600 ns and 3.4 µs (two components); not a canonical property name | dopant thallium, 0.10 to 0.12 mole%temperature not stated on the page | — |
| 21 | Decay time, slow component, CsI(Tl) | Printed on the BNC CsI(Tl) datasheet in the same decay time row as the 600 ns component; not a canonical property name | dopant thallium, 0.10 to 0.12 mole%temperature not stated on the page | — |
| 22 | Decay time, CsI(Na) | Printed on the Shalom EO CsI(Na) page as 630 ns; not a canonical property name | dopant sodiumtemperature not stated on the page | — |
| 23 | Photoelectron yield with a bialkali PMT, CsI(Tl) | Printed on the BNC CsI(Tl) datasheet as 45% relative to NaI(Tl); not a canonical property name | readout bialkali photomultiplier, gamma raysdopant thallium, 0.10 to 0.12 mole% | — |
| 24 | Photoelectron yield, CsI(Na) | Printed on the Shalom EO CsI(Na) page as 85% of NaI(Tl) for gamma rays; not a canonical property name | readout gamma raysdopant sodium | — |
| 25 | Lower wavelength cutoff, CsI(Tl) | Printed on the Shalom EO comparison chart; not a canonical property name | dopant thallium, 0.10 to 0.12 mole% | — |
| 26 | Refractive index at the emission maximum, undoped CsI | Printed on the Shalom EO comparison chart; the undoped value is at a different wavelength from the Crystran value at 10 µm, so the two are not compared as one property | wavelength 315 nm (emission maximum of undoped CsI)dopant none (undoped crystal) | — |
| 27 | Thallium doping concentration, CsI(Tl) | Printed on the BNC CsI(Tl) datasheet as 0.10 to 0.12 mole%; not a canonical property name | dopant thallium | — |
Corrosion and compatibility
What happens when it meets other materials and environments.
| Environment | Concern | What happens |
|---|---|---|
| Moist air | moderate | CsI is slightly hygroscopic, so bare crystals need a dry store.1 |
| Liquid water at 0 °C | severe | CsI dissolves in water, at 44 g per 100 g of water at 0 °C.2 |
| Low Earth orbit, trapped radiation | low | The light yield of the Fermi LAT CsI(Tl) calorimeter falls about 1 percent a year.3 |
| Thermal cycling | moderate | Large temperature ramps can crack CsI(Tl), because it expands more than most crystals.4 |
| # | Environment | Details | Test conditions |
|---|---|---|---|
| 1 | Moist air | The BNC datasheet calls CsI(Tl) slightly hygroscopic and gives a storage limit of less than 50 percent relative humidity for a bare crystal. The Shalom EO chart calls CsI(Na) hygroscopic, which is stronger. | basis qualitative statement and storage limit; no uptake rate printedagent water vapour |
| 2 | Liquid water at 0 °C | The Crystran table gives a solubility of 44 g per 100 g of water at 0 °C, which is far higher than the calcium fluoride figure. This is a solubility figure, not a corrosion rate, and no dissolution test is printed. | temperature 0 °Cnote solubility figure, not a corrosion rate |
| 3 | Low Earth orbit, trapped radiation | The LAT paper reports a slow decrease of about 1 percent a year in the scintillation light yield of the crystals from radiation damage in low Earth orbit. It fits a time constant of about 2 years and predicts an energy-scale shift of about 4.5 percent after 10 years, which the team plans to correct. | basis flight data over three years, and a fit with a time constant; not a ground testagent trapped charged particles in the South Atlantic Anomaly |
| 4 | Thermal cycling | BNC says CsI(Tl) has a relatively high thermal expansion coefficient and recommends slow ramp rates, and it gives a maximum thermal gradient of 10 °C per minute for small crystals. No cycle count is printed. | basis vendor guidance; no fatigue test printedagent temperature ramp |
Risks and side effects
Hazards to people, what happens when a cell overheats, and how the material is handled and stored.
For people
PubChem's aggregated GHS record for CsI lists H315, H317, H319 and H335. The percentages are notification shares, and no supplier SDS was read.
- Skin and eyes
- H315 (skin irritation), H317 (allergic skin reaction) and H319 (serious eye irritation), per PubChem's aggregated record.
- Inhalation
- H335 (respiratory irritation), per PubChem's aggregated record. The dopant thallium is toxic and is not sourced in this run (see gaps).
If it fails
The failure results come from vendor datasheets and one flight instrument. No flight failure record was read.
- Softness
- Crystran says the crystal is extremely soft, with an apparent elastic limit of 5.6 MPa, so it deforms and is hard to polish.
- Thermal stress
- BNC warns that the high thermal expansion can crack a crystal during fast temperature changes.
- Water
- Crystran gives a solubility of 44 g per 100 g of water at 0 °C, so a wet surface can dissolve the crystal.
Handling and storage
- Humidity
- Keep a bare crystal below 50 percent relative humidity, and use a sealed or lightly protected housing.
- Cleaning
- Do not clean or polish with water, because the crystal dissolves in water.
- Not covered
- BNC gives a storage limit of less than 50 percent relative humidity for a bare crystal. No storage temperature is printed.
Test details
The original wording and numbers from the sources, for readers who want them.
Failure risks
| Failure | Trigger and consequence |
|---|---|
| Plastic deformation under light loadmoderate severity | Mechanical contact, or polishing a windowThe Crystran table lists CsI as having no cleavage and as deforming, and it gives an apparent elastic limit of 5.6 MPa. The crystal therefore bends or scratches under light load, and Crystran says it is extremely difficult to polish. |
| Thermally induced stress fracturemoderate severity | Fast temperature ramp, especially in larger crystalsBNC warns that a high thermal expansion can cause stress fractures during temperature cycling and gives a maximum ramp of 10 °C per minute for small crystals. |
| Slow loss of light yield in orbitlow severity | Trapped radiation in low Earth orbitThe LAT calorimeter loses about 1 percent a year of light yield, which is corrected by a calibration applied channel by channel. The loss is slow and does not stop the instrument. |
| Dissolution in waterhigh severity | Liquid water or condensation on an uncoated crystalThe Crystran table gives a solubility of 44 g per 100 g of water at 0 °C, so a wet surface can dissolve the crystal. |
Handling and storage
- Store a bare crystal below 50 percent relative humidity, as the BNC datasheet advises.
- Use gentle handling and avoid polishing or cleaning with water, because the crystal is extremely soft and dissolves in water.
Mitigations
- Use a sealed or lightly protected housing for CsI(Tl). BNC says that a sealed or lightly protected housing lets the crystal tolerate normal laboratory humidity, and that hermetic encapsulation is available.
- Use an encapsulated crystal. Shalom EO says polished and encapsulated CsI(Na) crystals are available.
Supply and geography
How common the raw materials are and where they come from.
Caesium, the metal in the feedstock (USGS 2026 summary, data for 2025)
- Recycling
- USGS: cesium iodide is recycled from radiography panels and used in the production of new panels.
- Substitutes
- USGS: many applications have used mineral substitutes, because primary cesium is not available globally.
Iodine, the iodide part of CsI (world mine production, 2025)
- Top producers (2025)
- Chile 67.6%, Japan 26.5%
- Recycling
- USGS: small amounts of iodine were recycled.
US iodine import sources (2021 to 2024, as printed)
- Top producers (2024)
- Chile 88%, Japan 11%
- Recycling
- Not repeated here; see the first supply entry.
Documents
Datasheets, safety data sheets, papers and standards you can open.
Full record
The complete spore, as stored. Free to reuse under CC-BY-4.0. Images keep the licence shown with each one.
Show the full YAML
spore: profile
id: caesium-iodide
spore_version: 0.2
family: optical-crystals
name: Caesium iodide (CsI), with thallium and sodium doped scintillator grades
abbreviation: CsI
role: IR optical crystal, and X-ray and gamma scintillator (CsI(Tl) and CsI(Na))
category: glasses-optics
type: optical-crystals
summary: |
Caesium iodide is the cubic CsCl-structure crystal CsI, grown by the sealed ampoule Stockbarger method in ingots of about 70 mm diameter. Crystran gives a transmission range of 0.25 to 55 µm, a refractive index of 1.73916 at 10 µm and a density of 4.51 g/cc, and it calls CsI the material with the deepest known IR transmission. The crystal is very soft and hard to polish. Doped with thallium it becomes CsI(Tl), which emits at 550 nm and matches silicon photodiodes and SiPM arrays (Berkeley Nucleonics). Doped with sodium it becomes CsI(Na), which emits at 420 nm and matches bialkali photomultipliers (Shalom EO). The only flight record read is the Fermi LAT calorimeter, which uses CsI(Tl) in low Earth orbit and sees its light yield fall by about 1 percent a year.
best_for: IR windows that must reach far into the infrared, and X-ray and gamma scintillators read by silicon photodiodes or SiPMs (CsI(Tl))
weakest_at: 'Softness and moisture: the crystal is extremely soft and hard to polish, and it is slightly hygroscopic and soluble in water'
identity:
formula: CsI
cas: 7789-17-5
crystal_system: cubic
space_group: Pm3m, CsCl structure
source: crystran-csi-page
where: Product Data table, Class/Structure row (Cubic CsCl, Pm3m, no cleavage, deforms); molecular weight 259.83 on the Crystran table and 259.8099 for CID 24601 on PubChem; CAS 7789-17-5 from PubChem CID 24601
tags:
environments:
- low-earth-orbit
stresses: []
applications:
- optics-and-cameras
properties:
- benchmark: refractive_index
label: Refractive index at 10 µm
value: 1.73916
unit: '-'
scope: material
conditions:
wavelength: 10 µm
dopant: none (undoped crystal)
method: Printed on the Crystran product data table, with references to the Palik handbook and a 1955 paper
status: ground
source: crystran-csi-page
where: Product Data table, Refractive Index row (1.73916 at 10 µm)
- benchmark: refractive_index_temperature_coefficient
label: Refractive index change with temperature (dn/dT)
value: -99.3
unit: ppm/K
scope: material
conditions:
unit_in_source: 10^-6 K^-1
wavelength: not stated on the page
dopant: none (undoped crystal)
method: Printed as -99.3 x 10^-6 K^-1 with reference 2 (Rodney 1955); the number is not converted; not a canonical property name
status: ground
source: crystran-csi-page
where: Product Data table, dn/dT row
- benchmark: refractive_index
label: Refractive index at the emission maximum, CsI(Tl)
value: 1.79
unit: '-'
scope: material
conditions:
wavelength: 550 nm (emission maximum)
dopant: thallium, 0.10 to 0.12 mole%
method: Printed on the Berkeley Nucleonics CsI(Tl) datasheet (2016), specifications table
status: ground
source: bnc-csitl-datasheet
where: Specifications table, Refractive index at emission maximum row
- benchmark: refractive_index
label: Refractive index at the emission peak, CsI(Na)
value: 1.84
unit: '-'
scope: material
conditions:
wavelength: 420 nm (emission peak)
dopant: sodium
method: Printed on the Shalom EO CsI(Na) product page, specifications table and comparison chart
status: ground
source: shalomeo-csi-na
where: Specifications table, Refractive Index at Emission Peak row; comparison chart, Refractive Index at Emission Maximum row
- benchmark: density
label: Density
value: 4.51
unit: g/cc
scope: material
conditions:
temperature: not stated on the page
dopant: none (undoped crystal)
method: Value printed on the Crystran product data table; the same 4.51 g/cm3 is printed for CsI(Tl) and CsI(Na) on the BNC and Shalom EO pages
status: ground
source: crystran-csi-page
where: Product Data table, Density row
- benchmark: melting_point
label: Melting point
value: 621
unit: °C
scope: material
conditions:
pressure: not stated on the page
dopant: none (undoped crystal)
method: Value printed on the Crystran product data table; the Shalom EO comparison chart prints 621 °C for CsI, CsI(Tl) and CsI(Na)
status: ground
source: crystran-csi-page
where: Product Data table, Melting Point row
- benchmark: max_service_temp
label: Maximum useful temperature, CsI(Tl) crystal
value: 125
unit: °C
scope: material
conditions:
dopant: thallium, 0.10 to 0.12 mole%
basis: bare crystal, useful temperature range
method: Printed on the BNC CsI(Tl) datasheet as the upper end of a useful temperature range; the lower end is left out (see gaps)
status: ground
source: bnc-csitl-datasheet
where: Specifications table, Useful temperature range row
- benchmark: thermal_conductivity
label: Thermal conductivity
value: 1.1
unit: W/(m·K)
scope: material
conditions:
temperature: 298 K
dopant: none (undoped crystal)
method: Printed on the Crystran product data table with reference 3 (Combes et al. 1951); the BNC datasheet prints 0.13 for CsI(Tl), which is not reconciled here (see gaps)
status: ground
source: crystran-csi-page
where: Product Data table, Thermal Conductivity row
- benchmark: cte
label: Coefficient of thermal expansion
value: 48.3
unit: ppm/K
scope: material
conditions:
unit_in_source: 10^-6 K^-1
temperature: 293 K
dopant: none (undoped crystal)
method: Printed as 48.3 x 10^-6 K^-1 at 293 K; the number is not converted
status: ground
source: crystran-csi-page
where: Product Data table, Thermal Expansion row
- benchmark: cte
label: Coefficient of thermal expansion, CsI(Tl)
value: 51
unit: ppm/K
scope: material
conditions:
unit_in_source: 10^-6 K^-1
temperature: not stated on the page
dopant: thallium, 0.10 to 0.12 mole%
method: Printed on the BNC CsI(Tl) datasheet as 51 x 10^-6 K^-1; the number is not converted; the Shalom EO chart gives 54 for CsI(Tl) (see gaps)
status: ground
source: bnc-csitl-datasheet
where: Mechanical and Thermal Properties table, Thermal expansion coefficient row
- benchmark: cte
label: Coefficient of thermal expansion, CsI(Na)
value: 49
unit: ppm/K
scope: material
conditions:
unit_in_source: 10^-6 K^-1
temperature: not stated on the page
dopant: sodium
method: Printed on the Shalom EO CsI(Na) page as 49x10^-6 K^-1; the number is not converted
status: ground
source: shalomeo-csi-na
where: Specifications table, Thermal Expansion Coefficient row
- benchmark: specific_heat
label: Specific heat capacity
value: 201
unit: J/(kg·K)
scope: material
conditions:
temperature: not stated on the page
dopant: none (undoped crystal)
method: Printed on the Crystran product data table with reference 4 (Kelly 1934); the BNC datasheet prints the same 201 J/kg K
status: ground
source: crystran-csi-page
where: Product Data table, Specific Heat Capacity row
- benchmark: dielectric_constant
label: Dielectric constant at 1 MHz
value: 5.65
unit: '-'
scope: material
conditions:
frequency: 1 MHz
dopant: none (undoped crystal)
method: Value printed on the Crystran product data table
status: ground
source: crystran-csi-page
where: Product Data table, Dielectric Constant row
- benchmark: elastic_modulus
label: Young's modulus
value: 5.3
unit: GPa
scope: material
conditions:
temperature: not stated on the page
dopant: none (undoped crystal)
method: Value printed on the Crystran product data table; the BNC datasheet prints 5.3 GPa for CsI(Tl)
status: ground
source: crystran-csi-page
where: Product Data table, Youngs Modulus (E) row
- benchmark: shear_modulus
label: Shear modulus
value: 6.24
unit: GPa
scope: material
conditions:
temperature: not stated on the page
dopant: none (undoped crystal)
method: Value printed on the Crystran product data table
status: ground
source: crystran-csi-page
where: Product Data table, Shear Modulus (G) row
- benchmark: bulk_modulus
label: Bulk modulus
value: 12.67
unit: GPa
scope: material
conditions:
temperature: not stated on the page
dopant: none (undoped crystal)
method: Value printed on the Crystran product data table
status: ground
source: crystran-csi-page
where: Product Data table, Bulk Modulus (K) row
- benchmark: poissons_ratio
label: Poisson ratio
value: 0.214
unit: '-'
scope: material
conditions:
temperature: not stated on the page
dopant: none (undoped crystal)
method: Value printed on the Crystran product data table
status: ground
source: crystran-csi-page
where: Product Data table, Poisson Ratio row
- benchmark: apparent_elastic_limit
label: Apparent elastic limit
value: 5.6
unit: MPa
scope: material
conditions:
temperature: not stated on the page
dopant: none (undoped crystal)
method: Printed on the Crystran product data table as 5.6 MPa (810 psi); not a canonical property name
status: ground
source: crystran-csi-page
where: Product Data table, Apparent Elastic Limit row
- benchmark: reststrahlen_peak
label: Reststrahlen peak
value: 145.8
unit: µm
scope: material
conditions:
temperature: not stated on the page
dopant: none (undoped crystal)
method: Printed on the Crystran product data table; not a canonical property name
status: ground
source: crystran-csi-page
where: Product Data table, Reststrahlen Peak row
- benchmark: transmission_range
label: Transmission range
value: 0.25
value_high: 55
bound: range
unit: µm
scope: material
conditions:
thickness: not stated on the page
dopant: none (undoped crystal)
method: Range printed on the Crystran product data table; not a canonical property name
status: ground
source: crystran-csi-page
where: Product Data table, Transmission Range row
- benchmark: water_solubility
label: Solubility in water at 0 °C
value: 44
unit: g/100 g water
scope: material
conditions:
temperature: 0 °C
dopant: none (undoped crystal)
method: Printed on the Crystran product data table; not a canonical property name
status: ground
source: crystran-csi-page
where: Product Data table, Solubility row
- benchmark: emission_peak_wavelength
label: Emission maximum, CsI(Tl)
value: 550
unit: nm
scope: material
conditions:
dopant: thallium, 0.10 to 0.12 mole%
temperature: not stated on the page
method: Printed as the maximum emission wavelength on the BNC CsI(Tl) datasheet; the Shalom EO comparison chart also prints 550; not a canonical property name
status: ground
source: bnc-csitl-datasheet
where: Specifications table, Maximum emission wavelength row
- benchmark: emission_peak_wavelength
label: Emission peak, CsI(Na)
value: 420
unit: nm
scope: material
conditions:
dopant: sodium
temperature: not stated on the page
method: Printed on the Shalom EO CsI(Na) page as the emission peak wavelength; not a canonical property name
status: ground
source: shalomeo-csi-na
where: Specifications table, Emission Peak Wavelength row
- benchmark: decay_time
label: Decay time, fast component, CsI(Tl)
value: 600
unit: ns
scope: material
conditions:
dopant: thallium, 0.10 to 0.12 mole%
temperature: not stated on the page
method: Printed on the BNC CsI(Tl) datasheet as 600 ns and 3.4 µs (two components); not a canonical property name
status: ground
source: bnc-csitl-datasheet
where: Specifications table, Decay time row
- benchmark: decay_time_slow_component
label: Decay time, slow component, CsI(Tl)
value: 3.4
unit: µs
scope: material
conditions:
dopant: thallium, 0.10 to 0.12 mole%
temperature: not stated on the page
method: Printed on the BNC CsI(Tl) datasheet in the same decay time row as the 600 ns component; not a canonical property name
status: ground
source: bnc-csitl-datasheet
where: Specifications table, Decay time row
- benchmark: decay_time
label: Decay time, CsI(Na)
value: 630
unit: ns
scope: material
conditions:
dopant: sodium
temperature: not stated on the page
method: Printed on the Shalom EO CsI(Na) page as 630 ns; not a canonical property name
status: ground
source: shalomeo-csi-na
where: Specifications table, Decay Time row
- benchmark: photoelectron_yield_relative_nai_tl
label: Photoelectron yield with a bialkali PMT, CsI(Tl)
value: 45
unit: '% of NaI(Tl)'
scope: material
conditions:
readout: bialkali photomultiplier, gamma rays
dopant: thallium, 0.10 to 0.12 mole%
method: Printed on the BNC CsI(Tl) datasheet as 45% relative to NaI(Tl); not a canonical property name
status: ground
source: bnc-csitl-datasheet
where: Specifications table, Photoelectron yield row
- benchmark: photoelectron_yield_relative_nai_tl
label: Photoelectron yield, CsI(Na)
value: 85
unit: '% of NaI(Tl)'
scope: material
conditions:
readout: gamma rays
dopant: sodium
method: Printed on the Shalom EO CsI(Na) page as 85% of NaI(Tl) for gamma rays; not a canonical property name
status: ground
source: shalomeo-csi-na
where: Specifications table, Photoelectron Yield row
- benchmark: lower_wavelength_cutoff
label: Lower wavelength cutoff, CsI(Tl)
value: 320
unit: nm
scope: material
conditions:
dopant: thallium, 0.10 to 0.12 mole%
method: Printed on the Shalom EO comparison chart; not a canonical property name
status: ground
source: shalomeo-csi-na
where: Comparison chart, Lower Wavelength Cutoff row, CsI(Tl) column
- benchmark: refractive_index
label: Refractive index at the emission maximum, undoped CsI
value: 1.95
unit: '-'
scope: material
conditions:
wavelength: 315 nm (emission maximum of undoped CsI)
dopant: none (undoped crystal)
method: Printed on the Shalom EO comparison chart; the undoped value is at a different wavelength from the Crystran value at 10 µm, so the two are not compared as one property
status: ground
source: shalomeo-csi-na
where: Comparison chart, Refractive Index at Emission Maximum row, CsI (Undoped) column
- benchmark: dopant_concentration
label: Thallium doping concentration, CsI(Tl)
value: 0.1
value_high: 0.12
bound: range
unit: mole%
scope: material
conditions:
dopant: thallium
method: Printed on the BNC CsI(Tl) datasheet as 0.10 to 0.12 mole%; not a canonical property name
status: ground
source: bnc-csitl-datasheet
where: Specifications table, Thallium doping concentration row
corrosion:
- environment: Moist air
headline: CsI is slightly hygroscopic, so bare crystals need a dry store.
effect: The BNC datasheet calls CsI(Tl) slightly hygroscopic and gives a storage limit of less than 50 percent relative humidity for a bare crystal. The Shalom EO chart calls CsI(Na) hygroscopic, which is stronger.
rating: moderate
conditions:
basis: qualitative statement and storage limit; no uptake rate printed
agent: water vapour
source: bnc-csitl-datasheet
where: Specifications table, Hygroscopicity and Storage condition rows; Shalom EO comparison chart, Hygroscopic row
- environment: Liquid water at 0 °C
headline: CsI dissolves in water, at 44 g per 100 g of water at 0 °C.
effect: The Crystran table gives a solubility of 44 g per 100 g of water at 0 °C, which is far higher than the calcium fluoride figure. This is a solubility figure, not a corrosion rate, and no dissolution test is printed.
rating: severe
conditions:
temperature: 0 °C
note: solubility figure, not a corrosion rate
source: crystran-csi-page
where: Product Data table, Solubility row
- environment: Low Earth orbit, trapped radiation
headline: The light yield of the Fermi LAT CsI(Tl) calorimeter falls about 1 percent a year.
effect: The LAT paper reports a slow decrease of about 1 percent a year in the scintillation light yield of the crystals from radiation damage in low Earth orbit. It fits a time constant of about 2 years and predicts an energy-scale shift of about 4.5 percent after 10 years, which the team plans to correct.
rating: low
conditions:
basis: flight data over three years, and a fit with a time constant; not a ground test
agent: trapped charged particles in the South Atlantic Anomaly
source: arxiv-1206-1896
where: Calorimeter section, text on radiation damage in the first three years; section 7.3.2 fit and prediction
- environment: Thermal cycling
headline: Large temperature ramps can crack CsI(Tl), because it expands more than most crystals.
effect: BNC says CsI(Tl) has a relatively high thermal expansion coefficient and recommends slow ramp rates, and it gives a maximum thermal gradient of 10 °C per minute for small crystals. No cycle count is printed.
rating: moderate
conditions:
basis: vendor guidance; no fatigue test printed
agent: temperature ramp
source: bnc-csitl-datasheet
where: Mechanical and Thermal Properties table, Maximum thermal gradient rows; thermal note below the specifications
risks:
hazards:
- text: PubChem's aggregated GHS record for CsI gives H315 (causes skin irritation), H317 (may cause an allergic skin reaction), H319 (causes serious eye irritation) and H335 (may cause respiratory irritation). The percentages PubChem prints are shares of supplier notifications, not a classification by a regulator.
ghs:
- H315
- H317
- H319
- H335
source: pubchem-cid-24601
where: GHS Hazard Statements list, CID 24601 record
failure_risks:
- mode: Plastic deformation under light load
trigger: Mechanical contact, or polishing a window
consequence: The Crystran table lists CsI as having no cleavage and as deforming, and it gives an apparent elastic limit of 5.6 MPa. The crystal therefore bends or scratches under light load, and Crystran says it is extremely difficult to polish.
severity: moderate
source: crystran-csi-page
where: Class/Structure row; Apparent Elastic Limit row; Crystran product description
- mode: Thermally induced stress fracture
trigger: Fast temperature ramp, especially in larger crystals
consequence: BNC warns that a high thermal expansion can cause stress fractures during temperature cycling and gives a maximum ramp of 10 °C per minute for small crystals.
severity: moderate
source: bnc-csitl-datasheet
where: Thermal note below the specifications; Maximum thermal gradient rows
- mode: Slow loss of light yield in orbit
trigger: Trapped radiation in low Earth orbit
consequence: The LAT calorimeter loses about 1 percent a year of light yield, which is corrected by a calibration applied channel by channel. The loss is slow and does not stop the instrument.
severity: low
source: arxiv-1206-1896
where: Calorimeter section, text on radiation damage; section 7.3.2
- mode: Dissolution in water
trigger: Liquid water or condensation on an uncoated crystal
consequence: The Crystran table gives a solubility of 44 g per 100 g of water at 0 °C, so a wet surface can dissolve the crystal.
severity: high
source: crystran-csi-page
where: Product Data table, Solubility row
side_effects: []
handling:
- text: Store a bare crystal below 50 percent relative humidity, as the BNC datasheet advises.
source: bnc-csitl-datasheet
where: Specifications table, Storage condition row
- text: Use gentle handling and avoid polishing or cleaning with water, because the crystal is extremely soft and dissolves in water.
source: crystran-csi-page
where: Product Data table, Solubility row; product description
mitigations:
- text: Use a sealed or lightly protected housing for CsI(Tl). BNC says that a sealed or lightly protected housing lets the crystal tolerate normal laboratory humidity, and that hermetic encapsulation is available.
source: bnc-csitl-datasheet
where: Introduction text on hygroscopicity; Wrapping row of the configurations table
- text: Use an encapsulated crystal. Shalom EO says polished and encapsulated CsI(Na) crystals are available.
source: shalomeo-csi-na
where: Product description text
uses:
- application: optics-and-cameras
device: IR spectrophotometer components
example: Crystran says CsI is sometimes used for components in the widest range spectrophotometers, because it has the deepest known IR transmission.
source: crystran-csi-page
where: Product description, first paragraph
- application: optics-and-cameras
device: X-ray image intensifier input phosphor and X-ray panels
example: USGS says cesium iodide is used primarily in X-ray panel production, and in fluoroscopy equipment as the input phosphor of X-ray image intensifier tubes.
source: usgs-mcs2026-cesium
where: Domestic Production and Use paragraph
- application: optics-and-cameras
device: Gamma-ray telescope calorimeter (Fermi LAT)
example: The Fermi LAT calorimeter has 8.6 radiation lengths of CsI(Tl) crystals stacked in 8 layers. Its crystal logs measure 326 mm by 26.7 mm by 19.9 mm, and the paper reports the light yield loss in orbit.
source: arxiv-1206-1896
where: Introduction, paragraph describing the CAL; calorimeter module description
- application: optics-and-cameras
device: Silicon photodiode and SiPM readout of CsI(Tl) detectors
example: BNC says the 550 nm CsI(Tl) emission matches standard silicon photodiodes and SiPM arrays, so it can be read without a photomultiplier.
source: bnc-csitl-datasheet
where: Introduction text; Typical Applications list
supply:
- subject: Caesium, the metal in the feedstock (USGS 2026 summary, data for 2025)
producers: []
critical_lists: []
recycling: 'USGS: cesium iodide is recycled from radiography panels and used in the production of new panels.'
substitutes:
- 'USGS: many applications have used mineral substitutes, because primary cesium is not available globally.'
source: usgs-mcs2026-cesium
where: Domestic Production and Use paragraph; Recycling paragraph. The USGS summary gives no official world production figure and no country shares. It names Canada, China, Germany and Russia as the primary global producers, estimated without shares, in the Salient Statistics paragraph.
- subject: Iodine, the iodide part of CsI (world mine production, 2025)
producers:
- country: Chile
share_percent: 67.6
year: 2025
- country: Japan
share_percent: 26.5
year: 2025
critical_lists: []
recycling: 'USGS: small amounts of iodine were recycled.'
substitutes: []
source: usgs-mcs2026-iodine
where: 'World Mine Production and Reserves table, 2025 column: Chile 23,000 and Japan 9,000, of a world total of 34,000 in metric tons of elemental iodine. Shares calculated by us as country value divided by the printed world total. The United States is shown as W, withheld.'
- subject: US iodine import sources (2021 to 2024, as printed)
producers:
- country: Chile
share_percent: 88
year: 2024
- country: Japan
share_percent: 11
year: 2024
critical_lists: []
recycling: Not repeated here; see the first supply entry.
substitutes: []
source: usgs-mcs2026-iodine
where: Import Sources (2021 to 2024) line, printed as Chile 88 percent, Japan 11 and other 1. The year shown is the last year of the range.
documents:
- title: 'Crystran: Caesium Iodide CsI optical material page, Product Data table (undated)'
type: datasheet
url: https://www.crystran.com/optical-materials/caesium-iodide-csi
year: 2026
- title: 'Berkeley Nucleonics ScintIQ: CsI(Tl) Cesium Iodide Scintillation Crystals datasheet (revision 04-01-2016)'
type: datasheet
url: https://www.berkeleynucleonics.com/cesium-iodide/
year: 2016
- title: 'Hangzhou Shalom EO: CsI(Na) Scintillator Crystals product page, with CsI, CsI(Tl) and CsI(Na) comparison chart'
type: datasheet
url: https://www.shalomeo.com/Scintillators/CsI-Na/product-392.html
year: 2026
- title: 'PubChem: Cesium iodide (CID 24601), compound record with GHS classification'
type: database
url: https://pubchem.ncbi.nlm.nih.gov/compound/24601
year: 2026
- title: 'PubChem: Caesium (CID 5354618), compound record with molecular weight'
type: database
url: https://pubchem.ncbi.nlm.nih.gov/compound/5354618
year: 2026
- title: 'USGS Mineral Commodity Summaries 2026: Cesium'
type: report
url: https://pubs.usgs.gov/periodicals/mcs2026/mcs2026-cesium.pdf
year: 2026
- title: 'USGS Mineral Commodity Summaries 2026: Iodine'
type: report
url: https://pubs.usgs.gov/periodicals/mcs2026/mcs2026-iodine.pdf
year: 2026
- title: 'The Fermi Large Area Telescope On Orbit: Event Classification, Instrument Response Functions, and Calibration (arXiv:1206.1896)'
type: paper
url: https://arxiv.org/pdf/1206.1896
year: 2012
gaps:
- No ground or orbital radiation test of undoped CsI or CsI(Na) was found in a source we could open. The only in-orbit result is the Fermi LAT calorimeter light yield trend for CsI(Tl), which is an instrument measurement and not a material test.
- No outgassing (TML/CVCM) or atomic-oxygen result for caesium iodide was found in the sources read.
- Thermal conductivity conflicts between sources. Crystran gives 1.1 W/(m·K) at 298 K for undoped CsI; BNC gives 0.13 W/(m·K) for CsI(Tl). Both are entered with their own grade conditions and are not reconciled.
- Thermal expansion conflicts between sources. Crystran gives 48.3 ppm/K at 293 K for undoped CsI; BNC gives 51 for CsI(Tl); the Shalom EO chart gives 54 for CsI(Tl) and 49 for CsI(Na). The 54 is not entered, to keep one value per grade.
- Light output in photons per MeV is printed by Shalom EO as 52 to 56 thousand for CsI(Tl) and 38 to 44 thousand for CsI(Na). It is not entered as a property, because the printed form (x10^3) does not match the value check; the relative photoelectron yield is entered instead.
- 'Decay time for CsI(Tl) conflicts: BNC gives 600 ns and 3.4 µs, and the Shalom EO chart gives 1 µs. The BNC two-component figure is entered and the 1 µs is not.'
- The lower end of the useful temperature range for CsI(Tl) is printed as -40 °C with a typographic minus sign. It is left out of the numeric fields and only the upper end is entered.
- The Crystran data table prints Knoop hardness as Knoop 20 with a 200 g indenter, with no unit. It is not entered.
- The refractive index of undoped CsI at 10 µm (Crystran, 1.73916) and at 315 nm (Shalom EO, 1.95) are at different wavelengths. Both are entered with their own conditions and are not compared as one property.
- Thallium is toxic and is the dopant in CsI(Tl). No thallium safety data sheet or thallium GHS record was read in this run, so the dopant hazard is not sourced.
- No supplier safety data sheet for CsI was opened. The GHS entry is PubChem's aggregated notification record, which prints notification shares, not a regulator's classification. The Crystran page links a 'Caesium Bromide (CsI) SDS' under a name that does not match the grade, and it was not opened.
- The atomic weight of iodine was not read; only the molecular weight of I2 (253.8089, CID 807) was seen. Iodine is therefore listed in the recipe without an atomic weight.
- The USGS cesium summary gives no official world production figure and no country shares, so no caesium production shares are entered. The producer list it names is estimated, not measured.
- Supply shares for iodine are calculated by us from the printed world total and are shares of mine output, not of CsI crystal output.
- The Fermi LAT crystal count of 1536 appeared only in a search summary and is not entered.
- The vendor pages give values for their own grades, not independent measurements. The Crystran page is undated and the BNC datasheet is dated 2016.
- No media (photograph or drawing with a licence) was checked for this profile.
license: CC-BY-4.0
updated: '2026-10-26'
review:
status: draft
method: assistant
checked_values: 31
unreachable: 0
caveats: 5
note: 'Written by an AI assistant on 2026-10-26 from the Crystran product page, the Berkeley Nucleonics CsI(Tl) datasheet, the Shalom EO CsI(Na) page, PubChem records, the USGS 2026 cesium and iodine summaries and the Fermi LAT paper read in the same run. Caveats: the vendors conflict on thermal conductivity, thermal expansion and decay time; no supplier SDS was opened; the thallium dopant hazard is not sourced; caesium supply has no official figures; light output in photons per MeV is not entered.'
sources:
crystran-csi-page:
title: 'Crystran: Caesium Iodide CsI optical material page, Product Data table'
id: Crystran product page for caesium iodide (undated)
url: https://www.crystran.com/optical-materials/caesium-iodide-csi
year: 2026
kind: datasheet
license: supplier page, linked not reproduced
bnc-csitl-datasheet:
title: 'Berkeley Nucleonics ScintIQ: CsI(Tl) Cesium Iodide Scintillation Crystals datasheet'
id: CsI(Tl)_datasheet, revision 04-01-2016
url: https://www.berkeleynucleonics.com/cesium-iodide/
year: 2016
kind: datasheet
license: vendor datasheet, linked not reproduced
shalomeo-csi-na:
title: 'Hangzhou Shalom EO: CsI(Na) Scintillator Crystals product page, with comparison chart for CsI, CsI(Tl) and CsI(Na)'
id: product 392, CsI(Na) page
url: https://www.shalomeo.com/Scintillators/CsI-Na/product-392.html
year: 2026
kind: datasheet
license: vendor page, linked not reproduced
pubchem-cid-24601:
title: PubChem compound record, cesium iodide, CID 24601, with GHS classification
id: CID 24601; CsI; molecular weight 259.8099
url: https://pubchem.ncbi.nlm.nih.gov/compound/24601
year: 2026
kind: database
license: public domain (NCBI)
pubchem-cid-5354618:
title: PubChem compound record, caesium, CID 5354618, with molecular weight
id: CID 5354618; Cs; molecular weight 132.9054520
url: https://pubchem.ncbi.nlm.nih.gov/compound/5354618
year: 2026
kind: database
license: public domain (NCBI)
usgs-mcs2026-cesium:
title: Mineral Commodity Summaries 2026, Cesium (USGS, Candice C. Tuck)
id: USGS MCS 2026, data for 2025
url: https://pubs.usgs.gov/periodicals/mcs2026/mcs2026-cesium.pdf
year: 2026
kind: report
license: public domain (USGS)
usgs-mcs2026-iodine:
title: Mineral Commodity Summaries 2026, Iodine (USGS)
id: USGS MCS 2026, data for 2025
url: https://pubs.usgs.gov/periodicals/mcs2026/mcs2026-iodine.pdf
year: 2026
kind: report
license: public domain (USGS)
arxiv-1206-1896:
title: 'The Fermi Large Area Telescope On Orbit: Event Classification, Instrument Response Functions, and Calibration'
id: arXiv:1206.1896
url: https://arxiv.org/pdf/1206.1896
year: 2012
kind: paper
license: arXiv, linked not reproduced
story:
recipe:
line: Caesium iodide is an ionic compound of caesium and iodine, with one caesium atom for each iodide ion. Crystran grows the crystal by the sealed ampoule Stockbarger technique in ingots of about 70 mm diameter. Thallium or sodium is added as a dopant to make the scintillator grades, CsI(Tl) at 0.10 to 0.12 mole percent thallium and CsI(Na) with sodium.
parts:
- name: Caesium
symbol: Cs
atoms: 1
atomic_weight: 132.905452
role: The alkali metal cation. One caesium atom per formula unit (PubChem CID 5354618 for the atomic weight).
- name: Iodine
symbol: I
role: The halogen anion. One iodine atom per formula unit (formula CsI from PubChem CID 24601).
- name: Thallium
symbol: Tl
role: Activator dopant in the CsI(Tl) grade, at 0.10 to 0.12 mole percent (BNC).
- name: Sodium
symbol: Na
role: Activator dopant in the CsI(Na) grade (Shalom EO).
risk_groups:
- title: For people
level: 1
verdict: Irritant; may cause allergy or breathing irritation
lead: PubChem's aggregated GHS record for CsI lists H315, H317, H319 and H335. The percentages are notification shares, and no supplier SDS was read.
points:
- label: Skin and eyes
line: H315 (skin irritation), H317 (allergic skin reaction) and H319 (serious eye irritation), per PubChem's aggregated record.
- label: Inhalation
line: H335 (respiratory irritation), per PubChem's aggregated record. The dopant thallium is toxic and is not sourced in this run (see gaps).
- title: If it fails
lead: The failure results come from vendor datasheets and one flight instrument. No flight failure record was read.
points:
- label: Softness
line: Crystran says the crystal is extremely soft, with an apparent elastic limit of 5.6 MPa, so it deforms and is hard to polish.
- label: Thermal stress
line: BNC warns that the high thermal expansion can crack a crystal during fast temperature changes.
- label: Water
line: Crystran gives a solubility of 44 g per 100 g of water at 0 °C, so a wet surface can dissolve the crystal.
- title: Handling and storage
gap: BNC gives a storage limit of less than 50 percent relative humidity for a bare crystal. No storage temperature is printed.
points:
- label: Humidity
line: Keep a bare crystal below 50 percent relative humidity, and use a sealed or lightly protected housing.
- label: Cleaning
line: Do not clean or polish with water, because the crystal dissolves in water.
why: Caesium iodide is a cubic crystal that transmits from 0.25 to 55 µm (Crystran), so it is used where the infrared reaches furthest. Doped with thallium it becomes a 550 nm scintillator that silicon photodiodes read (BNC). Doped with sodium it becomes a 420 nm scintillator that bialkali photomultipliers read (Shalom EO). The Fermi LAT calorimeter flew CsI(Tl) in low Earth orbit and measured a light yield loss of about 1 percent a year.
uses:
- where: IR spectrophotometers
line: Crystran says CsI is sometimes used for components in the widest range spectrophotometers, because it has the deepest known IR transmission.
- where: X-ray panels and image intensifiers
line: USGS says CsI is used primarily in X-ray panel production, and as the input phosphor of X-ray image intensifier tubes in fluoroscopy.
- where: Gamma-ray telescope (Fermi LAT)
line: The LAT calorimeter has 8.6 radiation lengths of CsI(Tl) crystals in 8 layers, and its light yield is tracked in orbit.
can_do:
- label: Range
value: 0.25 to 55 µm
line: Crystran's data table gives a transmission range of 0.25 to 55 µm for undoped CsI, the widest of the windows listed on the page.
- label: Refraction
value: '1.73916'
line: Refractive index 1.73916 at 10 µm, per the Crystran data table. Refractive index 1.79 at the CsI(Tl) emission maximum (BNC).
- label: Light
value: 550 nm
line: CsI(Tl) emits at 550 nm (BNC) and CsI(Na) emits at 420 nm (Shalom EO), so each grade suits a different photosensor.
- label: Heat
value: 621 °C
line: Melting point 621 °C on the Crystran table. BNC gives a useful temperature range up to 125 °C for CsI(Tl).
- label: Density
value: 4.51 g/cc
line: Density 4.51 g/cc for the crystal, the same value printed for CsI(Tl) and CsI(Na).
compare: Crystran's sheet and the two scintillator vendors give different figures for each grade, so the comparison is between grades and not between suppliers. Undoped CsI transmits to 55 µm, and its thermal conductivity is 1.1 W/(m·K) at 298 K on the Crystran table, a value higher than the 0.13 that BNC gives for CsI(Tl). CsI(Tl) is about 51 ppm/K in thermal expansion on the BNC sheet, and CsI(Na) is 49 on the Shalom EO chart. The sheets give no side-by-side ground or space test and no cost comparison.
compare_rows:
- label: Heat conduction
benchmark: thermal_conductivity
- label: Thermal expansion
benchmark: cte
- label: Stiffness
benchmark: elastic_modulus
handling:
- Keep bare crystals below 50 percent relative humidity, as BNC advises.
- Do not clean or polish with water, because CsI dissolves in water (44 g per 100 g at 0 °C).
- Handle gently and avoid pressure on the surface, because the crystal is extremely soft.
good_for:
- IR windows and spectrophotometer components that must transmit to 55 µm (Crystran).
- X-ray panels and image intensifier input phosphors (USGS).
- CsI(Tl) scintillators read by silicon photodiodes or SiPMs (BNC).
watch_out:
- Humidity. CsI is slightly hygroscopic (CsI(Tl)) and hygroscopic (CsI(Na) per Shalom EO). Keep bare crystals dry.
- Softness. The crystal deforms and is hard to polish, so the transmission quality is compromised for range.
- Thermal ramps. The high thermal expansion means slow ramp rates are needed for larger crystals (BNC).
- Orbit. The Fermi LAT calorimeter lost about 1 percent a year of light yield, and a correction is applied.
- Source conflicts. Vendors disagree on thermal conductivity, thermal expansion and decay time (see gaps).
verdict: Caesium iodide is a soft, moisture-sensitive crystal with the widest IR window of the common halides, and as CsI(Tl) it is a rugged scintillator that a silicon photodiode can read. The sources read describe its optics and its one flight use, but give no material test in space, so its orbital life is not established by this profile.
References
Every number above carries a reference number. Click one to see the source and where in it the value comes from.
| # | Source | Type | Year | Used for |
|---|---|---|---|---|
| Crystran: Caesium Iodide CsI optical material page, Product Data table | datasheet | 2026 | 22 values | |
| Berkeley Nucleonics ScintIQ: CsI(Tl) Cesium Iodide Scintillation Crystals datasheet | datasheet | 2016 | 14 values | |
| Hangzhou Shalom EO: CsI(Na) Scintillator Crystals product page, with comparison chart for CsI, CsI(Tl) and CsI(Na) | datasheet | 2026 | 8 values | |
| PubChem compound record, cesium iodide, CID 24601, with GHS classification | database | 2026 | 1 value | |
| PubChem compound record, caesium, CID 5354618, with molecular weight | database | 2026 | 0 values | |
| Mineral Commodity Summaries 2026, Cesium (USGS, Candice C. Tuck) | report | 2026 | 2 values | |
| Mineral Commodity Summaries 2026, Iodine (USGS) | report | 2026 | 2 values | |
| The Fermi Large Area Telescope On Orbit: Event Classification, Instrument Response Functions, and Calibration | paper | 2012 | 3 values |